GCE Medical products now found on gce-medical.com EXPLORE NOW close
Acasa » MAXIFLOW 2/2 Acetylene

MAXIFLOW 2/2 Acetylene

Acetylene manifold | two inlets with semi-automatic changeover

Part. No.
Gas Description Pressure Inlet Outlet
F27310035 A with automatic quick-acting shut-off device according to ISO 14114-2018 25/1,5 bar W21,8×1/14“LH male G1/2” male
F27310036 A

with automatic quick-acting shut-off device according to ISO 14114-2018

and REED contact gauge

25/1,5 bar W21,8×1/14“LH male G1/2” male
0768166 A with manual quick-acting shut-off device according to ISO 14114-2014 25/1,5 bar W21,8×1/14“LH male G1/2” male
0768167 A

with manual quick-acting shut-off device according to ISO 14114-2014

and REED contact gauge
25/1,5 bar W21,8×1/14“LH male G1/2” male


TECHNICAL DATA

Body material: Brass
Connectors
& fittings material:
Brass
Diaphragm material: EPDM
Seat sealing material: Chloroprene
Wall bracket: Stainless steel
Maximal inlet pressure: 25 bar
Maximal flow rate: 5 Nm3/h
Temperature range: from -20°C to 60 °C

 For more details please refer to the data sheet.

Technical datasheets

Maxiflow-2/2-acetylene-EN

Informatii tehnice Sus

Descarcati informatiile tehnice

Acrobat Logo DATA-TEC Maxiflow 2-2 Acetylene-EN.pdf